VI. BATTERY CHANGE
When the SPY RF battery has to be replaced, the LCD screen displays the following message:
To replace the battery, follow the instructions below:
1
Never unscrew.
This will cancel the
warranty
BEWARE: KEEP THE BATTERY FAR FROM FIRE. DON'T TRY TO RELOAD IT OR TO SHORT-
CIRCUIT ITONLY USE BATTERIES SUPPLIED BY JRI (REF: 06569)
VII. RESET
If the device does not work anymore (cannot turn it on...), use the Reset function in the same way as the battery
change.
VIII. FEATURES
FEATURES
Measurement range
Number of channels
Type of input
Accuracy
Operating conditions
Temperature for storage
Memory size
Recording interval
Radio range (in free field)
Radio band
Battery lifetime
Dimensions
Protection level
CE ERM conformity
Conformity FCC
DOWNLOAD THE MEMORY BEFORE CHANGING THE
BATTERY. NEVER WAIT UNTIL THE BATTERY IS
EMPTY OR THE DATA WILL BE DELETED.
2
1
3
2
a) Unplug the battery connector.
b) Short circuit the 2 pins with a screwdriver
Open the
c) Connect the new battery
battery cover
d) Change the seal
e) Close the battery cover
±0,4°C from -20 to +30°C and ±2% HR from 20 to 80%HR
, ±3.5% HR from 10 to 20% and from 80 to 90% HR
18
3
a
SPY RF TH
-30°C +70°C and 0 to 100%HR
not condensed
2
Built in TH digital sensor
±0,5°C out side and
-30°C +70°C
-40°C +85°C
1s to 90 min
20 000 measurements
1 km
868MHz or 902 MHz
2 years
123x69x30mm
IP34
EN 301 489 / EN 61000 / EN 61010
EN 55022 / EN 300 220
FCC part 15
b