*THESE CHECKS SHOULD INCLUDE:
17. LIFTING STRAP: Your lift strap should be checked at service intervals
for any sign of wear (ideally every six months). In order to view the
strap it is recommended to remove the cap at the top of the mast, cycle
the lift in both directions whilst observing the condition of the strap. If
you see any wear (fraying, creasing or similar) it is recommended an
authorised, Joerns Healthcare Dealer or qualified, competent service
engineer replaces the strap IMMEDIATELY. For details regarding strap
maintenance and replacement please request our detailed service guide.
(See below for details)
B. 1 Periodic inspection of the hoist should be undertaken at the time intervals stated by the
manufacturer, but at least once a year. By periodic inspection is meant a visual examination
(particularly of the hoist's load bearing structure and lifting mechanism with attachments, brakes,
controls, safety devices and person-support devices) and whatever function tests and maintenance
measures may be required, e.g. adjustment of brakes, tightening of fasteners.
Every inspection should include a working load test of one (1) lifting cycle with the maximum load.
22
Check
Lift Strap
EN ISO 10535:2006
(Informative)
Periodic Inspection
Annex B
ü
Check
Lift Strap
English
ü

Publicité