9 Error messages
Error 1 Pressure sensor failure or overpressure.
Error 2 PT100 probe failure or overtemp.
Error 3 External alarm. Pressure switch or safety thermostat trigered.
To reset the safety thermostat press its button number
23
The thermostat is adjusted in the factory it must not be ope-
rated except to reset it.
Error 4 Manual valve (16) opened during a cicle.
Error 5 Opened door during a cycle.
Error 6 Pressure over +0.05bar with the autoclave stopped.
Error 7 E2PROM failure.
Error 8 Out of limits temperature during the sterilization.
Error 9 Mains supply failure during cycle.
10 Spare parts
Presoclave II 80 litres. Part No. 4002136
07292
RS-232 communication wire
07315
26 pins communication wire
13006
Contactor
15411
Fuse 16 A
16055
Solenoid electromagnet
16057
Electrovalve NA 3 mm 24 V
16212
Pressure switch
20002
Lid microrruptor
20067
Valve microrruptor
20096
Bipolar switch
29358
Frontal circuit
29362
Power circuit
36027
Solid relay 25A
39001
Heating element 3680W/230V
43031
Thermostat
43071
Pt100 probe
Presoclave II 150 litres. Part No. 4002137