INSTRUCTION MANUAL CODE 80219 REV E
24
15
16
J
.P. SELECTA s.a.u.
7 Installation
7.1 Locate the autoclave close to an appropriate power source.
7.2 The autoclave needs to be placed on a flat stable horizontal sur-
face, leaving a minimum space of 10 cm around the instrument.
7.3 Adjust the front feet to level the autoclave and set the autoclave
in position.
7.4 Fix a plastic or rubber hose to the pump output connection (24)
with a receptical to collect any condensate produced by the at-
mospheric purge.
¡WARNING! DON'T TRY TO OPEN THE LID UN-
LESS READING THE «OPERATION» SECTION
8 Operation
8.1 Opening the Lid
-
The autoclave must be connected to mains supply and the main
switch turned on.
-
Pressure gauge: 0 bar
-
Temperature in the internal chamber must be under 95ºC.
-
The steam manual valve (16) must be opened.
ONLY WITH THIS CONDITIONS LID IS READY TO OPEN
8.2 Start
-
Start on using the main switch (18)
-
Open the lid as we can see on 8.1 section, using the handle (17)
-
Place the shelf at the bottom of chamber.
-
Be sure the valve (15) is closed.
17
-
Fill of decalcified water until shelf level.
-
Put the material to sterilize and close the lid.
-
Close the manual emptying valve (16)
-
Start the cycle as shows the section 8.3