6 Equipment description
The Presoclave II autoclave belongs to a series of saturated vapour autoclaves that
are distinguished for the following features:
1. 6 point radial locking system.
2. Temperature and time electronic control circuit.
3. Safety blocking lock that only allows the opening of the lid with pressure 0 and
temperature under 95ºC.
4. Maximum pressure limiter that disconnects the heating element in cases of over-
pressure.
5. Limiter thermostat to protect the heating element.
6.1 Equipment
-
Safety system:
Safety overpressure valve.
Safety pressure switch.
Safety thermostat.
Lid closed detector.
Protective thermal lid cover.
-
Chamber and lid made from stainless steel.
-
External case made from stainless steel.
-
Manual release vapour valve.
-
Manual emptying valve.
-
Waste water tank connection.
6.2 Features
-
Sterilization temperature controllable from 115°C to 134°C.
-
Temperature resolution 1ºC.
-
Sterilization time from 3 minutes to 99 minutes.
-
Time resolution 1 minute.
-
Suitable to work with solids and liquids atmospheric purge from 115°C to
134°C.