cleaning and disinfecting
CLEANING AND DISINFECTION OF MEDICINE CONTAINER
1. Open the lid of the container after each use and pour out the remaining liquid.
2. Pour a small amount of hot water into the container (water temperature 40ºC -50ºC),
close the lid, shake it and empty it;
3. Pour 2ml of hot water again, press and hold the power button for 5 seconds to enter
cleaning mode (intermittent misting, red light on). Once the hot water runs out,
press the power button to turn off and wipe off the remaining medicine with clean
medical gauze.
4. Rinse the inside and outside walls of the container and lid with pure water. After
cleaning, use a clean medical gauze to gently wipe away the water. The vibrating
mesh can be cleaned with a medical cotton swab and the action should be as gentle
as possible.
5. The medicine container can be immersed in 75% medical alcohol for at least 3
minutes for disinfection, then drain it and place it in a clean place to dry.
REMOVE THE MEDICINE CONTAINER
Press the release button down and, at the same time, pull the container backwards to
separate it from the main unit.
NOTE
1. During the cleaning and disinfection process it is necessary to use clean and
sterilized medical gauze and medical cotton swabs.
2. Do not use paper towels or other cloths to dry the water, otherwise dust or lint
of cloth may remain on the parts, preventing the device from working.
3. If the metal mesh plate is blocked (low or no atomization), pour 2 ml of hot
water, max 60°C, into the medicine container, press the power button for 5
seconds to enter cleaning mode and atomize until 2 ml of liquid is used up.
4. Turn off the device then open the lid of the medicine container and immerse it
in warm water or a solution of water and vinegar for about 3 minutes, then rinse
with pure water.
5. Do not touch the mesh with sharp and/or hard objects.
6. Do not use detergents other than those recommended. Turn off the power and
unplug the power cord when cleaning and disinfecting.
7. Always open the lid of the medicine container when you immerse it in water or
alcohol to clean it.
HOW TO CLEAN THE MAIN UNIT
1. If there are stains on the outer surface of the main unit, moisten a gauze with 75%
medical alcohol, wipe it gently, and then dry it with dry gauze.
2. A dry gauze can be used to clean the electrodes on the main unit and the medicine
container to ensure that the electrodes are connected, and the device can work
normally.
30