USAGE
The nebuliser is designed for respiratory diseases treatment and respiratory tract
infection prevention. It is widely used in curing diseases of respiratory system, such as
cold, fever, bronchitis.
APPLICABILITY
The nebulizer can be used by adults, the elderly, infants and children under the
supervision of an adult. The nebuliser can be used at home-hold environment, and it
complies with requirement of IEC 60601-1-11.
Contraindications: 1) The device cannot be applied to anaesthesia breathing system and
respiratory system. 2) Infants and children must under adults' supervision. 3) Pentamidine
is not applicable to the device.
WORKING PRINCIPLE
The nebulisation occurs through the high-speed vibration of the mesh plate which
vaporizes the medicinal liquid that comes out through the micro mesh holes into very
small particles.
1. Atomized particles
2. Piezoelectric ceramic plate
3. Mesh plate
4. Liquid medicine
5. Mesh micropore
MAIN PERFORMANCE INDEX
Please see Table, it was tested by 0.9% physiological saline under the condition of
temperature 25 ± 3ºC, humidity 60% ± 5%R.H.
Working frequency
Nebulisation rate
Medicinal liquid temperature
Noise level
Particle size
1
MAIN PERFORMANCE INDEX
25
2
1 to 3 MHz
≥0.2ml/min
20°C to 40°C
≤25dB
5 μm ±25%
3
4
5