Test Procedure
1. Ensure the power on the equipment under test is OFF
before beginning.
2. Confirm the VOLTAGE ADJUST Knob is set to the zero-voltage
position, (fully counter clockwise).
3. Connect all supply leads or terminals on the item under test
according to the specifications for this test.
4. Turn on all power switches, if applicable, on the item under test.
5. Attach the black-wire ground clip to an exposed metal part of
the item under test.
WARNING: Ensure the alligator clip is connected to ground ONLY! It
must be connected first and removed last.
Never hold it by hand while applying high voltage to the unit under test!
6. Turn on the test set power switch. Both the green POWER
Indicator Light and the amber LINE FAULT Indicator Light
will illuminate.
7. Momentarily press the RESET button. The LINE FAULT Indicator
Light will turn OFF, and the red High Voltage Indicator Light
will illuminate.
8. Touch the high-voltage test probe to the terminal being tested
(e.g., AC line input prongs on a typical small appliance). Extend
the probe by pushing the plastic screw-head on the side.
9. Turn the VOLTAGE ADJUST Knob clockwise until the voltmeter
reads the desired test voltage, and maintain this voltage for the
specified time.
10. The leakage current will be indicated on the milliammeter.
11. Return the voltage to zero and remove the probe.