Measurement and instrumentation
Dual trace oscilloscope JO 20 MHz
Ref :
291166
ENGLISH
All manuals and user guides at all-guides.com
2.3
Quick start – Rear panel
(31) Z INPUT
This input is used to modulate the trace brightness via an external signal.
A (+) signal reduces the brightness; a (-) signal increases it.
(32) YA OUTPUT
Replica of the YA 20 mV/div. input signal, e.g. for feeding into a frequency
meter.
(33) FUSE HOLDER / VOLTAGE SELECTOR (see 2.1. Installation)
(34) POWER SOCKET
3
Operation – Instructions sheet (see
attachment)
Instructions Sheet: Oscilloscope
Use the oscilloscope in display and voltmeter mode.
Shunt connection
Before: All pushbuttons in up position (pulled out)
•
SCREEN BRIGHTNESS and ILLUMINATION low
•
VERTICAL and HORIZ POSITION at mid range
•
TRIGGERING on AUTO and INT
•
VARIABLE knobs in CAL position
Oscilloscope power-up
Settings:
•
reasonable BRIGHTNESS
•
SHARPNESS on the finest trace
•
VERTICAL YA lever on O, adjust POSITION (trace on axis)
•
HORIZ POSITION as required
How to set the oscilloscope
1st question
What do we want to
display and
measure?
22
1 voltage
YA = f(t)
YA = f(t)
and YB = f(t)
Press on YA
Press on YA and YB
or
2 voltages
YB = f(YA)
XY
Set YB
Set X and Y
BRIGHTNES
S