Silicate HR PP / M352
Silicate HR PP
1 - 90 mg/L SiO
Silicomolybdate
Material
EN
Required material (partly optional):
Reagents
VARIO Silica HR Reagent, Set F10
Preparation
1.
The temperature of the sample should be between 15 °C and 25 °C.
Notes
1.
The method measures in the flank of the absorption curve of the resulting
coloration. For filter photometers, the accuracy of the method can therefore be
improved, if necessary, by user adjustment using a silicate standard (approx. 70
mg/L SiO
2
18
2
).
Packaging Unit
1 Set
EN   Method Reference Book   1.0
M352
SiHr
Part Number
535700