Silicate LR PP / M351
EN
Swirl around to dissolve the
powder.
Blank
Place blank in the sample
chamber. Pay attention to
the positioning.
Swirl around to dissolve the
powder.
Once the reaction period is finished, the measurement takes place automatically.
14
Press the ENTER button.
Sample
Add a Vario Silica Amino
Acid F10 powder pack to
the sample vial.
Zero
Press the ZERO button.
EN   Method Reference Book   1.0
1 min
12
11
1
10
2
9
3
8
4
7
5
6
Wait for 1 minute(s)
reaction time.
Close vial(s).
2 min
12
11
1
10
2
9
3
8
4
7
5
6
Wait for 2 minute(s)
reaction time.