What to do, if ...
... the data logger cannot be contacted
from the interface:
- Check that the interface has been
connected correctly and that the
evaluation
recognized it.
- Open the data logger (see also the
pictures in the "Replacing the
Battery" section).
- Check
to
condensation has penetrated the
data logger housing.
- Either wipe the condensation from
them housing or allow it to air dry
by leaving the data logger housing
open.
- Pull the battery plug from the
socket and wait 30 seconds.
- Then reconnect the battery and
check that the LED to the right of
the
battery
flashing. If the LED does not flash,
the battery may be flat and needs
to be replaced (see "Replacing the
Battery")
- As a test, insert a fresh battery in
the socket on the circuit board.
- If the LED still does not flash
(despite the fresh battery), the
problem
lies
logger.
- Before you close the data logger
housing again, replace the silicone
O-ring.
ba80000def01_EBI16
software
has
see
whether
compartment
is
with
the
data
Français
Que faire si ...
... l'enregistreur
communication avec l'interface :
- Vérifier
que
correctement connectée et que le
logiciel d'évaluation l'a reconnue.
- Ouvrir
l'enregistreur
schema
„remplacement de la pile").
- Vérifier si de la condensation s'est
infiltrée
dans
l'enregistreur.
- Essuyer
le
condensation du boîtier ou laisser
sécher à l'air libre en laissant le
boîtier ouvert.
- Débrancher les connectiques de la
pile et attendre 30 secondes.
- Ensuite reconnecter la pile et
vérifier que la led à droite de la
pile clignote. Si la led ne clignote
pas, la pile est peut être vide et a
besoin d'être remlacée (voir la
rubrique
„remplacement
pile).
- Pour effectuer un test, insérer une
autre pile neuve.
- Si la led ne clignote toujours
pas, le problème provient de
l'enregistreur lui-même.
- Avant de fermer l'enregistreur,
remplacer le joint d'étanchéité en
silicone.
n'entre
pas
l'interface
a
(voir
de
la
rubrique
le
boîtier
cas
échéant
de
en
été
les
de
la
la
27