4.3 Installation and commissioning
1. Fit the EasyClean to a wall or a pipe/post, so that the connections are facing downwards.
2. Connect the transmitter to the EasyClean device (see section 4.2 "Connections").
3. Connect compressed air and water from the building, taking care to ensure that the media
pressure does not exceed the specified device pressure. (see section 1.3 "Data and key
characteristics")
4. Connect hoses to the compressed air/water output and install the corresponding housing
connections.
5. Connect the main cable.
Terminal for
N and L wires
Cable sleeve
6. Fitting the spray head for the EasyClean100:
b
d
c
a
When connecting the grounding wire use the
supplied blade terminals.
Crimp the terminals from both sides.
a Push the spray head over the end
of the InDip housing that is to be
immersed. The distance between
the air outlet opening and the elec-
trode should be approx. 1 cm.
b Install the hose system as follows:
Place a screw in the holding band
on the InDip housing, and use this
screw to secure the bracket to the
upper end of the housing.
c Insert the hose between the spray
head and the bracket as follows:
First of all, loosen the hose screw.
Push the hose by hand, without the
use of any tools, into the screw
housing, tighten the screw.
d Secure the loose hose to the tube
of the InDip housing using a cable
binder.
EasyClean100/150