Step
Settings
Display
1.
O/P off
Cal count
2.
0V
Cal count
3.
Max V, 100mA
Cal count
4.
Max V, 100mA
O/P volts
5.
I pot at 0
I pot
6.
2V, 0mA
Cal count
7.
2V, 0mA
O/P amps
8.
2V, max I
Cal count
9.
2V, max I
O/P amps
10.
2V, 0mA
Cal count
11.
2V, 0mA
O/P amps
12.
2V, 500mA
Cal count
13.
2V, 500mA
O/P amps
14.
O/P off
'End'
The Manufacturers or their agents overseas will provide a repair service for any unit developing a
fault. Where owners wish to undertake their own maintenance work, this should only be done by
skilled personnel in conjunction with the service manual which may be purchased directly from
the Manufacturers or their agents overseas.
Cleaning
If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild
detergent.
WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER
ALLOW WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER
CLEAN WITH SOLVENTS.
52
Adjust for
Load
0V ± 2mV
DVM
0V ± 2mV
DVM
15/30/60V ± 2mV
DVM
Display = DVM
DVM
Display = 0
0mA ± 0.2mA
milli-ammeter
Display = milli-ammeter
milli-ammeter
Change load
1.5/3A ± 1mA
Ammeter
Display = ammeter
Ammeter
0mA ± 0.02mA
milli-ammeter
Display = milli-ammeter
milli-ammeter
500mA ± 0.1mA
ammeter
Display = ammeter
ammeter
-
Detail
O/P off volts
V DAC zero
V DAC scale
V metering scale
-
I pot ADC zero
High I DAC zero
High I metering zero
High I DAC scale
High I metering scale
Low I DAC zero
Low I metering zero
Low I DAC scale
Low I metering scale
-
Stores and exits
Maintenance