6
Operation Method
6.1 Preparation before use
1) Please read the instruction manual thoroughly.
2) Check the appearance of the product for abnormalities such as deformation or damage.
3) Check the battery charge.
4) Inspect the handpiece and components for defects.
6.2 Instruction for use and operation procedures
1) Connect the adapter and power cord, insert the plug into the outlet.
2) Connect the DC plug to the charger.
3) Charging starts when the handpiece is placed in the charger. Indicator lamp will light up in
orange when charging, and green when charging is complete.
4) After removing the handpiece from the charger, secure the ultrasonic tip to the handpiece using
the provided wrench. Tighten until the ultrasonic tip no longer spins. Do not over-tighten.
5) Press and hold the Operation Button for about 1 second to turn on the power. When the button is
pressed in the standby mode, it turns on immediately.
6) Check that the Battery Level Indicator LED light is green.
7) Fill pulp chamber with NaOCI, EDTA, or other final rinse solution and insert the ultrasonic tip.
8) When the Operation Button is pressed, the operation status LED light turns on and operate for 15
seconds.
9) After operation, when the Operation Button is pressed or after 15 seconds, the operation status
LED turns off and operation stops.
10) Charge when the orange LED on the battery level indicator lights up. Charge immediately when
the orange LED flashes.
11) After use, remove the ultrasonic tip and clean it. For cleaning instructions, refer to 8. Cleaning,
Sterilization and Maintenance.
For safety, do not use continuously, follow the specified duty cycle(refer to 13. Specifications).
[ Adapter, Charger Connection]
10
6.3 Ultrasonic Tip Installation / Removal
Never screw or unscrew ultrasonic tip while the unit is activated.
1
Hand thread tip onto the handpiece. Insert the wrench into the
1
ultrasonic tip joint.
2
Turn (clockwise) within the range of the wrench's maximum operating
2
angle (about 25°). The wrench is designed to move only within a maximum
of about 25°. Force should only be applied within the designed angle.
Do not overtighten.
3
When the force is released, wrench will automatically revert to its
3
original position.
* Turn counterclockwise for ultrasonic tip removal.
6.4 Technique Instructions
Shape the root canal to be cleaned.
1
1
Fill the pulp cavity with cleaning agent / irrigation solutions
2
(NaOCl, EDTA, etc.).
Insert the ultrasonic tip into the handpiece.
3
Tighten the ultrasonic tip with the provided wrench.
4
Press the start button to start root canal irrigation.
5
2
Move ultrasonic tip up and down using a small (2-3mm) vertical motion,
6
maintaining a distance of 2mm from working length.
Activate the irrigation solution with the device 2-3 times for optimal result.
7
Suction out any remaining debris from the canals.
8
Avoid using the device in severely curved canals due to the risk of
6
ultrasonic tip fracture.
Avoid hitting the ultrasonic tip against the root canal wall.
3.0mm
2.0mm
1.0mm
11