5.3
Error Messages
This information can be used to identify and troubleshoot any error
messages that may appear in the load display (Figure 7-1, Item 2) of the
Wet Tensile Strength Tester.
If an error message appears in the load display, it is necessary to press
the Clear Tensile Display button (Figure 7-1, Item 7) to recover the
operating system of the Wet Tensile Strength Tester.
The following list shows all potential error messages that may occur in
the load display. The list provides basic definitions and potential causes
and solutions for each error message of the Wet Tensile Strength Tester:
"Er.1" - Capacity of load display exceeded. This message
appears in the load display if the tensile load required to break the
sand specimen is greater than the capacity of the instrument. If this
message appears, the current test is automatically and immediately
aborted.
"Er.2" - Test not ready to start. This message indicates that all
required start-up conditions were not achieved prior to pressing the
start button. Check the three possible reasons for this error
message:
»
Sand specimen heating time was not preset.
»
The local display was not zeroed.
»
The Ready to Start indicator (Figure 7-1, Item 3) was off.
Once the reason for the error message is identified, make the required
corrections and press the Start key to restart the test.
"Er.4" - Minimum load range exceeded. This message indicates
that negative loads are detected by the electronic load cell. This
may indicate that part of the measuring fork (Figure 7-2, Item 2) has
been removed or an upward force is or has been exerted on the
measuring fork. If the status of the measuring fork appears normal,
then this message may indicate a failure in the electronic circuit.
Contact Simpson Technologies for additional assistance in resolving
this message.
man-STC-42112-V17-3 French