7 Specifications
Recommended Settings
System pressure:
Spray air:
Spray water:
Ambient Conditions of Control Unit and Power Supply
Location:
Ambient temperature:
Relative humidity:
Max. altitude:
Storage and Transport Conditions of Control Unit and Power Supply
Danger when start-up the medical device after storage in very cold conditions.
This can cause an operational failure of the medical device.
Very cold devices must be brought to a temperature of 20 to 25° C (68 to 77° F) before being started up.
Ambient temperature:
Relative humidity:
Air pressure:
Keep dry!
We reserve the right to make technical modifications.
Type Label
The explanations of the symbols are described in chapter 2 Important Information.
Page 20
3.0 (43.5) bar (psi)
1.0 (14.5) bar (psi)
0.8 (11.6) bar (psi)
Permitted in interior rooms
10 to 40 (50 to 104) ° C (° F)
30 to 75 %
2,000 m
-30 to 70 (-22 to 158) ° C (° F)
5 to 95 %
700 to 1,060 hPa
E-STATIS Prep 40-P