7.4

Elektromagnetische en magnetische immuniteit

Immunity test
Electrostatic discharges
(IEC61000-4-2)
Electrical fast Transients
IEC61000-4-4)
Surge
(IEC61000-4-5)
Magnetic field at power
frequency
(IEC-61000-4-8)
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
(IEC61000-4-11)
Voltage interruptions
(IEC61000-4-11)
IEC60601 test level
± 8Kv contact
± 15Kv air
± 2Kv for power supply lines ± 2Kv for power supply lines
± 1Kv line(s) to line(s)
± 2Kv line(s) to earth
30A/m
0% UT for 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°.
0% UT for 1 cycle
70% UT
For 25 cycles at 50Hz
For 30 cycles at 60Hz
Monophasé à 0°
0% UT
For 250 cycles at 50Hz
For 300 cycles at 60Hz
Compliance level
± 8Kv contact
± 15Kv air
± 1Kv for input/output lines
± 1Kv line(s) to line(s)
± 2Kv line(s) to earth
30A/m
0% UT for 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°.
0% UT for 1 cycle
70% UT
For 25 cycles at 50Hz
For 30 cycles at 60Hz
Monophasé à 0°
0% UT
For 250 cycles at 50Hz
For 300 cycles at 60Hz
Electromagnetic environ-
ment
Home health environment
and a
professional health esta-
blishment
Home health environment
and a
professional health esta-
blishment
Home health environment
and a
professional health esta-
blishment
Home health environment
and a
professional health esta-
blishment
Home health environment
and a
professional health esta-
blishment
Home health environment
and a
professional health esta-
blishment