Warning!
Do not open lid until rotor has completely stopped! Avoid severe
personal injury or property damage from moving parts.
Rotor Replacement
To remove SnapSpin rotor:
-Press center button firmly
and at the same time lift rotor
upwards.
To install SnapSpin rotor:
-Set rotor onto the hex shaft.
Press center button firmly and
downward. Rotor should latch
with a click sound.
Warning!
Do not use damaged or cracked rotors!
Maintenance
Sprout
Plus is normally maintenance-free. Clean unit only when it is not
®
plugged into an electrical outlet. When necessary, the housing and rotor
can be wiped using a damp cloth and a mild, non-corrosive detergent.
Warning!
Do not use any solvent on the unit that may attack plastic or cause cracks
in the rotors or the unit. Avoid exposure to UV radiation, either natural
or man-made. In addition to cracks; discoloration, softening, or crazing
of the plastic materials can occur. Corrosion can also occur to the metal
parts which are part of the SnapSpinTM rotor feature. This damage
can weaken them. Ensure the rotor is protected from corrosion and
mechanical damage. The rotor must be cleaned with a neutral cleaning
liquid.
Notes
—Ensure all parts are thoroughly dry prior to operation.