Specifications
Environmental Ratings:
Note: for Indoor use
Altitude: up to 2 000m
Temperature: 2°C to 40°C
Maximum Relative Humidity: 80% for temperatures up to 31°C
decreasing linearly to 50% relative humidity at 40°C
Technical Features:
Capacity: (Max. liquid density: 1.2g/cm3 or 1.2g/mL)
Standard round 6-positon tube rotor: 12 mL Max
Strip tube (2x 8-position) rotor: 8 mL Max
Dimensions:
153mm x 128mm x 104mm (LxWxH)
Weight:
1.6 lb (0.73 kg) (Sprout
1.9 lb (0.86kg) (Sprout
Maximum Speed & RCF: (with fully loaded rotor)
6,000 rpm Approx., Max RCF: 2,000 x g
Electrical:
Power Adapter rating
Input:
Output: 12VDC, 2.2A
Sprout
Plus:
®
Input: 12VDC
Current (under load): Approx. 0.6A
Power: Approx. 7.3W
Accessories: (All accessories are rated for Maximum speed)
Tube adapters: 6x for 0.2mL tubes and 6x for
One SnapSpin strip tube rotor (HEA 100012)
(can be used in place of installed round rotor, HEA 100011)
This mark is the confirmation that the unit conforms to
the EU guidelines and has been tested according to the
following EU Declaration of Conformity.
Conforms to:
EN/IEC 61010-1 2010 ed.3
EN/IEC 61010-2-020 ed.3, 2016
ANSI/UL Std. 61010-1 ed.3, 2012
ANSI/UL Std. 61010-2-020 ed.3, 2016
710213
CAN/CSA C22.2 No.61010-1-12 (R2012)
CAN/CSA C22.2 No.61010-2-020 (R2017)
This product has been tested to the requirements of CAN/CSA C22.2 No.
61010-1 second edition, including Amendment 1, or a later version of the
same standard incorporating the same level of testing requirements.
Plus with rotor)
®
Plus with power adapter)
®
100-240VAC, 50/60Hz, 1.0A
0.5mL tubes (120350)