Phonomonon
Basic EMC standard
Immunity test levels
Electrostatic Discharge
Radiated RF EM field
Proximity fields from RF wireless
communicalions equipment
Hated power frequency
magnetic fields
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
l[C 61000-4-8
Home Healthcare Environment
±8 kV contact
, 2kV, , 4kV, , 8kV,
1
15kV air
10V/m 80MI lz-2.7GI lz 80% AM at 1kl lz
Refer to table 3
30Nm
50H7 or 60H7
Table 3 - Proximity fields from RF wireless communications equipment
Test frequency
Band
Immunity test levels
(MHz)
(MHz)
Professional healthcare facility environment
385
380-390
Pulse modulation 18Hz, 27V/rn
450
430-470
FM, ±5kHz deviation, 1 kHz sine, 28V/m
710
704-787
Pulse modulation 217Hz, 9V/m
745
780
810
800-960
Pulse modulation 18Hz, 28V/m
870
930
1720
Pulse modulalion 217HL, 28V/m
1845
1700-1990
1970
2450
2400-2570
Pulse modulation 217Hz, 28V/m
21