Technical Data
1. Power Supply
Battery operation: 6 AA size alkaline batteries, all connected in series
Battery operating time: Approximately 2 hours with 6 alkaline batteries
AC Adaptor operation: 110Vac / 220Vac wall plug type AC adaptor, galvanically
isolated with 9Vdc output voltage
Pump power dissipation:10.0 Watts maximum
Voltage regulation: Linear voltage regulator with low dropout voltage and short circuit
protection
2. Control System
Control unit: Microcontroller with integral program memory
Program memory: 32K flash ROM, 2K RAM memory
HMI: LCD display with LED backlight, soft keypad
Single / double control: Single or double breast pumping possible
Selectable pumping rhythm mode: 3 pumping rhythm modes are selectable and
controlled from keypad
Adjustable suction level: 8 suction levels are selectable and controlled from keypad
Suction vacuum level: Minimum 80 mmHg, maximum 220 mmHg
Technical Data
Power Requirements
AC Adapter (included)
AC Adapter – North American
Plug Configuration
110Vac/220Vac wall plug type
AC adaptor, galvanically isolated
adaptor with 9Vdc output voltage.
Batteries (not included)
6 AA batteries, 1.5V alkaline
Regulatory Compliance
UL (E302267 ) and CSA approved
with North American power adapter.
Troubleshooting
Troubleshooting the Lansinoh
®
Interruption of Electrical Service
If you experience an interruption in the normal operation of the electrical and/or
mechanical functioning of the pump, simply do the following:
Unplug the AC adapter cord from the electrical outlet.
n
Wait a few seconds, and then plug the AC adapter cord back into the electrical outlet.
n
If you still experience electrical failure, try using the pump with the six AA batteries
making sure the AC adapter is removed.
Discomfort While Pumping
If you feel any discomfort, you may need a different size flange.
Visit www.lansinoh.com/store or call 800-292-4794.
18
All manuals and user guides at all-guides.com
Regulatory Compliance
Type BF Applied Part
+122 ˚F
-4 ˚F
!
+122 ˚F
-4 ˚F
20% - 95%
20% - 80%
Caution
CLASS II
Transport /
Transport /
Operation
equipment
Storage
Storage
To maintain your power adaptor follow these instructions:
CORRECT
DO NOT unplug
DO NOT wrap cord
of power adaptor
power adaptor by
pulling on the cord.
around the plug body.
use power adaptor if damaged.
DO NOT
!
Warning: Children have STRANGLED in cords.
Keep the cords out of the reach of children.
Signature Pro
™
Double Electric Breast Pump
Decreased Suction
If it feels like the suction has decreased, try the following:
Remove the tubing from pump and place a finger over the tubing connector. If you
n
feel suction, the pump is operating properly, but the Breastmilk Expression Set may
not be correctly assembled or the white valve may be damaged. If no suction is felt
on your finger, try unplugging the adapter from the electrical outlet, wait a few
seconds, and plug it back into the electrical outlet.
Make sure the Breastmilk Expression Set has been properly assembled. Check the
n
assembly information in the section "Assembling the Breastmilk Expression Set",
Inspect the white valve for a tear or pinhole. This component is critical to achieving
n
proper suction. Replace the white valve with one of the spare valves provided with
the product.
When single pumping, check the tubing connector on the front of the pump to make
n
sure it is closed for single pumping. If open, close the connector.
If double pumping, ensure that the breastmilk expression set is on both breasts
n
before starting the pump.
Increased suction level for single pumping:
If it feels like the suction is higher than usual when single pumping, check that the plug
for the pipe is clean and that the airway in the plug is not blocked.
Operating
instructions
Limited Consumer Product Warranty (United States)
104 ˚F
This product is warranted to the original consumer purchaser (the "purchaser")
1008
41 ˚F
Operation
to be free from defects in material and workmanship which are not commercially
acceptable for the period of one year from the date of purchase on the pump
mechanism and the period of 90 days on all other product components. Warranty
coverage terminates if you sell or otherwise transfer this product to another person.
This warranty gives you specific legal rights and you may also have other rights,
CORRECT
which vary, by location.
LANSINOH
THE PRODUCT OTHER THAN THOSE WARRANTIES EXPLICITLY SET FORTH
HEREIN. ANY IMPLIED WARRANTY, INCLUDING ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR ANY IMPLIED WARRANTY OF FITNESS FOR A
PARTICULAR PURPOSE, TO THE EXTENT PERMITTED BY LAW, SHALL BE
LIMITED IN DURATION TO A PERIOD OF 90 DAYS FROM THE DATE OF
PURCHASE BY THE ORIGINAL PURCHASER.
In the event the product is found by Lansinoh
warranty, as purchaser's sole and exclusive remedy Lansinoh
option of Lansinoh
parts or labor.
NOTE: The purchaser shall bear all expense for returning this product to Lansinoh
This warranty does not apply to any product that has been subjected to misuse,
abuse, or alteration.
Only Lansinoh
breastfeeding products. Improper or incorrectly performed maintenance or repair,
or use of non-Lansinoh
Troubleshooting
Warranty
®
MAKES NO EXPRESS WARRANTY OF ANY KIND REGARDING
, replace this product without charge for such replacement or
®
brand replacement parts are recommended for use with Lansinoh
®
brand replacement parts may invalidate this warranty.
®
page 10
to not meet the above limited
®
will repair or, at the
®
.
.
®
®
19