The AS-200 suction equipment does not need maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. With regard to training, given the information contained in the user
manual and since it is easy to understand the said device, it doesn't appear to be necessary.
Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been damaged during
previous use. Connect the cable to electrical network and turn the switch on.
Close the aspirator outlet with your finger and with suction regulator at maximum check that the vacuum indicators reaches at
least -75kPa (-0.75 bar). Rotate the knob from right to left. The vacuum indicator should go down -25kPa (-0.25 bar). Check that no
loud noises are present. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and situated in the plug protects the
instrument. For fuses replacing, always the type and the range.
Before changing the fuse, disconnect the plug from the power supply socket.
Type of fault
1. The suction unit doesn't
work
2. No aspiration
3. No aspiration
4. The Vacuum power on the
patient side is either very low
or absent
5. The float doesn't close
6. The float doesn't close
7. Low suction
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7
If the overfill security system it's activated, don't proceede with the liquid aspiration.
If the overfill security system doesn't work there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device, in this case return the
device to CA-MI technical service.
CA-MI Srl will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information
that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
CA-MI TECHNICAL SERVICE. CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC
DIRECTIVE 2012/19/UE-WEEE:
The symbol on the device indicates the separated collection of electric and electronic equipment. At the end of
life of the device, don't dispose it as mixed solid municipal waste, but dispose it referring to a specific
collection centre located in your area or returning it to the distributor, when buying a new device of the
sample type to be used with the same functions.
This procedure of separated collection of electric and electronic devices is carried out forecasting a European
environmental policy aiming at safeguarding, protecting and improving environment quality, as well as avoiding potential effects
on human health due to the presence of hazardous substances in such equipment or to an improper use of the same or of parts of
the same Caution: The wrong disposal of electric and electronic equipment may involve sanctions.
PERIODICAL MAINTENANCE CHECKS
Cause
Cable is damaged
External power source failure
Jar Cap not screwed on correctly
Lid seal not in its seat
 Vacuum regulator set to
minimum
 Protection filter blocked or
damaged
 Connection tubes blocked, kinked
or disconnected
 Shut-off valve blocked or
damaged
 Pump motor damaged
If the cap has been washed, ensure
that the float is not partially
detached
The float it's covered by dirty
material
Foam inside the jar
None of the procedures have
achieved the desired results
Remedy
Replace the cable
Check the external power source
Unscrew the cap, and re-screw it
Unscrew the cap and insert the seal properly in
its seat
 Turn the vacuum regulator clockwise and
check the value of the vacuum on the gauge
 Replace the filter
 Replace or reconnect the tubes, check the jar
connections
 Empty the jar, or disconnect the tube from the
jar and unblock the shut-off valve. The unit
twill only work in the upright position
 Refer to authorised service personnel
Fit the float into it's place
Unscrewed the cap, leave the and put in on
autoclave
Fill the jar to 1/3 full of ordinary water
Contact CA-MI customer service