Audible Alarm Setting
Press the SET button five times and the audible alarm indicator
will
appear on the display. Press the UP or DOWN buttons to turn the audible
alarm On/Off. The instrument will exit the setup after 10 seconds of inac-
tivity or if the trigger is pulled at any time.
Trigger Lock Setup
Press the SET button six times and the temperature measurement trigger
will appear on the display. Press the UP or DOWN
lock indicator
buttons to turn the trigger lock setup on/off. When the temperature
measurement trigger lock is turned on, press the
button to enter the
time stamp function and the hour field will flash. Press the UP or DOWN
buttons to adjust the hour values. Press the
button again to switch to
the minutes field. To adjust, press the UP or DOWN buttons. Press the
a third time to confirm the entered time stamp. The instrument will exit the
setup after 10 seconds of inactivity or if the trigger is pulled at any time.
Turning High and Low Alarms ON/OFF
Press the HI/LO button to enable or disable the set alarms. To set alarms
refer to Setting the High Temperature Alarm or Setting the Low Tempera-
ture Alarm. When the high alarm limit is turned on as indicated by HI
on the display, the red LED and "HI" indicator will flash when the set
temperature is exceeded. If the audible alarm function has been turned
on, the meter will also emit a beeping sound.
When the low alarm limit is turned on as indicated by LO on the display,
the blue LED and "LO" indicator will flash when the measured value is
below the set alarm. If the audible alarm function has been turned on, the
meter will also emit a beeping sound.
continued...
13
www.REEDINSTRUMENTS.com