Object
& Root
:
Assembly
. Bur
│ ├
.Rates
│ │ ├
.Forward
│ │ │ ├
.Select
│ │ │ ├
.Digital
│ │ ├
.Reverse
│ │ │ ├
.Select
│ │ │ ├
.Digital
│ ├
.Fill
│ ├
.ModeDis
│ │ ├
.Select
│ │ ├
.V
│ │ ├
.Time
│ │ ├
.VStop
│ │ ├
.AutoFill
.Meas
│ ├
.Status
│ ├
.MeasInput
│ ├
.Ipol
│ ├
.Upol
.Outputs
│ ├
.AutoEOD
│ ├
.SetLines
│ │ ├
.L0
│ │ ├
up to L13
│ ├
.ResetLines
.Stirrer
798 MPT Titrino
&Assembly
Description
Assembly control
Buret
Rates
Forward rate
Type of rate control
Digital rate
as for forward rate
Type of rate control
Digital rate
Fill
Dispensing
Type of dispensing control
Volume to be dispensed
Time to dispense
Limit volume
Filling after each increment
Measuring
Measuring ON/OFF
Selection of measuring input
Polarization current
Polarization voltage
I/O outputs
Automatic output of EOD
Set I/O lines
Signal on L0
Reset I/O lines
Stirrer
3.2 Remote control commands
Input range
digital, analog
0...150, max.
digital, analog
0...150, max.
$G,$H,$C
$G,$S,$H,$C
volume, time
0.0001...0.1...9999
0.25...1...86 400
0.0001...9999, OFF
ON, OFF
ON, OFF
1, 2, diff., Ipol, Upol, Temp
0...1...±127
0...400...±1270
ON, OFF
$G
active,inactive,pulse,OFF
$G
ON, OFF
Reference
3.2.2.109.
ditto
ditto
ditto
3.2.2.110.
3.2.2.111.
ditto
ditto
ditto
ditto
ditto
3.2.2.112.
ditto
ditto
ditto
3.2.2.113.
ditto
ditto
ditto
3.2.2.114.