TROUBLESHOOTING
Problem
The power indicator (refer to
4.1) is not lit.
The power indicator is lit, but
Xenon lamp will not ignite.
12. CHART OF MEDICAL DEVICE SYMBOLS USED
LIT037Sunoptics Surgical
(English)
Solution
A. Check that the AC power cord is properly
connected.
B. Check the fuses. If necessary, replace.
A. Check to make sure lamp cartridge is seated properly.
B. Check that the lamp door is secured.
C. Check hour meter; if lamp hours exceed the rated lamp life,
replace the lamp (refer to 9.1).