9. MAINTENANCE, SERVICING AND REPAIR, WARRANTY
Performance of preventive maintenance is not essential. Regular maintenance can, however,
contribute to identifying potential problems before they become serious; thus, enhancing the
instrument's reliability and extending its useful operating life. Maintenance services can be
obtained from your local representative or from the manufacturer.
Defective items of equipment are to be serviced and repaired exclusively by persons
authorized by the manufacturer. All repair work shall employ original manufacturer's parts only.
9.1 LAMP REPLACEMENT
CERMAX lamps are at high internal pressure when cold and at
operating temperature; therefore, CERMAX lamps may unexpectedly
rupture resulting in discharge of hot fragments of quartz and/or glass
and metal.
Only handle lamps with protective covers in place. Do not handle lamps
without their protective covers unless government approved safety
glasses, facemasks (with neck protector, chest protector and
gauntlets) are worn.
Always disconnect power cord and turn main switch off before lamp
replacement.
CAUTION
Take utmost care not to hit gas-seal edge of lamp; it may cause gas
lamp failure.
CAUTION
Do not touch the lamp and patient simultaneously when changing the
lamp.
CAUTION
Open lamp compartment door and pull out lamp cartridge.
Replace lamp cartridge with new unit and log the serial number and hours on the hour
meter. Close the lamp compartment door. Re-connect the power cord and turn the light
source on according to 6.1. Reset hour meter to zero by depressing hour meter reset
button on rear panel.