Technical Data
Guidance and manufacturer's declaration-electromagnetic immunity
The MBPN002 and MB0500300 are intended for use in the electromagnetic environment specified below.
The customer or the user of the MBPN002 and MB0500300 should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency(50,
60 Hz)
magnetic field IEC
61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
IEC 60601
Compliance level
test level
+ 6 kV contact
+ 6 kV contact
+ 8 kV air
+ 8 kV air
+ 2kV for power
+ 2kV for power
supply lines
supply lines
+ 1kV for input/output
Not applicable
lines
+ 1kV line(s) to line(s)
+ 1kV differential
+ 2kV line(s) to earth
mode
Not applicable
<5% UT(>95% dip in
<5% UT(>95% dip in
UT) for 0,5 cycle
UT) for 0,5 cycle
40% UT(60% dip in
40% UT(60% dip in
UT) for 5 cycles
UT) for 5 cycles
70% UT(30% dip in
70% UT(30% dip in
UT) for 25 cycles
UT) for 25 cycles
<5% UT(>95% dip in
<5% UT(>95% dip in
UT) for 5 sec.
UT) for 5 sec.
3 A/m
3 A/m
Electromagnetic
environment-guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Main power quality should be
that of a typical commercial or
hospital environment.
Main power quality should be
that of a typical commercial or
hospital environment.
Main power quality should be
that of a typical commercial
or hospital environment. If
the user of the MBPN002
and MB0500300 requires
continued operation during
power mains interruptions,
it is recommended that the
MBPN002 and MB0500300 be
powered from an uninterruptible
power supply or a battery.
The MBPN002 and B0500300
power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
18