Cleaning and Disinfection
If the device is not cleaned and disinfected correctly and frequently as
indicated, microorganisms may remain in the unit and cause risk of infection.
After every use:
Clean the medication cup with distilled water after each inhalation.
1. Pour out the residual solutions in the medication cup.
2. Pour some distilled water into the medication cup.
3. Turn on the device to nebulize the distilled water for 1 to 2 minutes to clean the
mesh.
4. Remove batteries or disconnect the adapter from the main unit.
5. Remove the medication cup from the main unit.
6. Wash and rinse the medication cup with distilled water.
7. Shake off excess water and allow parts to be fully air dried on a clean, dry towel.
8. Use gauze or clean towel to wipe off stains on the main unit if necessary.
9. Make sure that all cleaned parts are completely dry before you store them or use
them next time.
1. Keep the battery compartment stay in dry all the time.
2. Do NOT poke the mesh with finger, cotton swab or any objects.
3. Do NOT clean parts in a dishwasher.
4. Do NOT use microwave to dry any parts.
Daily disinfections
It is important to disinfect the medication cup on a daily basis.
1. Disinfection by boiled water
- Rinse the medication cup with distilled water.
- Bring a saucepan of DISTILLED water to the boil.
- Carefully immerse the medication cup in the boiled water for a maximum ten (10) minutes.
- Carefully to remove the medication cup from the boiled water and shake off excess water.
- Allow parts cooling down and fully air dry on a clean, dry towel, and out of reach of
children.
- Make sure that all parts are clean and dry before storage or use.
2. Disinfection by alcohol
- Rinse the medication cup with distilled water.
- Immerse the medication cup in 75% ethyl alcohol for one (1) minute.
- Rinse the medication cup with distilled water again, shake off excess water and allow parts to
be fully air dried while lying on a clean dry towel.
- Make sure that all parts are clean and dry before storage or use.
12