Product HANDBOOK
Special Instructions for medical applications (IEC/EN/UL/CSA 60601-1)
These products are designed for continuous operation within an overall enclosure, and must be mounted
such that access to the mains terminals is restricted. See clause 16, IEC60601-1, EN60601-1 and UL/CSA
60601-1.
Input Connection – KMS15A–xx
Pin
Single O/P
1
2
3
+DC out
4
- DC out
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air or with
oxygen, or with nitrous oxide.
These products are NOT protected against the ingress of water.
Reference should be made to local regulations concerning the disposal of these products at the end of their
useful life.
Environmental Specifications:
Description
Use
Temperature
Case Temperature
Humidity
Altitude
Pressure
Orientation
Material Group
Pollution Degree
Overvoltage Category
Construction Class
Page 13 of 15
KMS15A Series
AC (L)
AC (N)
Operation
-40° C - +85° C
<115VAC is 83° C, 116VAC to 264VAC is 90° C
10 - 95% RH, non-condensing
70kPa - 106kPa
See mounting positions below
Indoor
5000m
IIIb
2
II
II
KMS15A Series 260258 issue 1, March 2016
Storage
indoor
-40° C - +90° C
-
10 - 95% RH, non-
condensing
5000m
54kPa - 106kPa
any