Rice Lake Weighing Systems OS-10 Guide De L'opérateur page 8

Publicité

Les langues disponibles

Les langues disponibles

Rice Lake OS-10/OS-25/OS-45
Calibration
Calibrate using the following process:
1. Ensure the device is powered off.
2. Press and hold
, ZERO displays.
3. Press
POWER ZERO
4. Press
UNITS
Units
5. Load appropriate amount of weight on scale (See
, CAL displays, then the weight.
6. Press
POWER ZERO
7. Remove the weight, re-power the scale and calibration is complete.
Error Messages
Low Battery
Over Load - Total load exceeds the
maximum capacity
Over Zero - Zero count is higher than
the setting range
Below Zero
Zero count is below the setting range
EEPROM error
8
Visit our website
while pressing
POWER ZERO
to select calibration unit.
Model
OS-10
OS-25
OS-45
Table 2. Calibration Weights
Error
Table 3. Error Messages
www.RiceLake.com
three times, ZERO COUNT displays.
UNITS
Units
Table
kg
10 kg
20 lb
20 kg
40 lb
20 kg
40 lb
Action
Replace the batteries
Remove weight from the scale
immediately
Scale turns off if status continues
30 secs or longer
Scale turns off if status continues
30 secs or longer
Call for service
2). Display shows span count.
lb
Display
Lo
Err
ErrL
ErrE