IDOL Manual
Model
Tolerance (±)
IDOL-100
IDOL-1KG
VIII.
Adjustment
1. Press and hold the MODE key until the display
shows "CAL"
2. Make sure nothing is on the scale platform then
press the MODE key again.
3. The display will flash "CAL" followed by the required
calibration weight (see chart above).
4. Place the required calibration weight on the platform.
The display will flash "PASS" then return to normal
weighing mode. Adjustment is complete!
IX. Troubleshooting
•
Problem: Scale will not turn on
•
Solution: Check the batteries and battery contacts.
•
Problem: Displayed weight fluctuates randomly
Solution: Make sure your workspace is stable and
•
free of any air currents or vibration. In some cases,
nearby radio wave emitting devices can cause
fluctuations.
•
Problem: Displayed weight is inaccurate
•
Solution: Perform a calibration adjustment. If
you are still having trouble, contact our warranty
department.
All manuals and user guides at all-guides.com
- English
Acceptable
Calibration
0.02g
0.2g
Recommended
Weight
Accuracy Class
100g
OIML M2
500g
OIML M2