IDOL Manual
Modèle
acceptable (±)
IDOL-100
IDOL-1KG
XX. Adjustment
1. Press and hold the MODE key until the display shows "CAL"
2. Make sure nothing is on the scale platform then press the
MODE key again.
3. The display will flash "CAL" followed by the required calibration
weight (see chart above).
4. Place the required calibration weight on the platform. The
display will flash "PASS" then return to normal weighing mode.
Adjustment is complete!
XXI.
Troubleshooting
Problem: Scale will not turn on
•
•
Solution: Check the batteries and battery contacts.
•
Problem: Displayed weight fluctuates randomly
•
Solution: Make sure your workspace is stable and
any air currents or vibration. In some cases, nearby radio wave
emitting devices can cause fluctuations.
•
Problem: Displayed weight is inaccurate
Solution: Perform a calibration adjustment. If you are still
•
having trouble, contact our warranty
XXII. Error Codes
1. OUTZ - Zero point has drifted out of the acceptable
range. Try to perform a calibration adjustment. If the
problem persists, the load cell may be permanently
All manuals and user guides at all-guides.com
- Français
Tolérance
Poids de
calibrage
0.02g
0.2g
Classe de
précision
recommandée
100g
OIML M2
500g
OIML M2
department.
free of