Section 3 Operation
Models GT1306 and GT1307
Figure 3-2. Reading Edge of Floats.
3-2
Figure 3-1. Typical Bypass Installation.
HORIZONTAL
LINE
FLOWMETER
A
E
A - Inlet Valve
D - Control Valve
Failure to drain the flowmeter when isolated in a bypass loop may
result in tube breakage caused by thermal expansion of the process
liquid.
Rate of flow is indicated by reading the increments inscribed on the
metering tube or direct etched scale parallel with the metering edge of the
float. For the correct reading edge of the float, refer to Figure 3-2.
Installation and Operation Manual
Part Number: 541B079AAG
VERTICAL
LINE
FLOWMETER
B
D
C
B - Outlet Valve
E - Drain Valve
CAUTION
X-VA-1306-1307-eng
October, 2016
D
B
C
A
E
C - Bypass Valve