9. BLUETOOTH TRANSMISSION
To transmit your measurement(s) onto on your smartphone or tablet, open the application
MyGluco and open the measurement screen. Turn on the Bluetooth function on your
smartphone or tablet and activate bluetooth mode on the glucose monitor.
There are two ways to activate the Bluetooth mode of the MyGluco and transmit the data to
your app. After the end of the measurement (the result is displayed on the screen) :
1. Keep the strip in the slot, and press the M button on the bottom left.
2. Eject the strip by pushing up the Strip-Ejection button or remove the strip directly by hand.
After following this step (1 or 2), "OFF" is shown and the Bluetooth mode is activated :
The blue LED (top right) will begin to blink, once the connection between the meter and
device is established, the blue LED will stop blinking and stays on permanently. The data is
transmitted to the app.
10. MAINTENANCE
10.1 BATTERY
Your meter comes with two 1.5V AAA size alkaline batteries.
LOW BATTERY SIGNAL
The meter will display one of the messages below to alert you when the meter power is
getting low.
1. The "
" symbol appears along with display messages: The meter is functional and the
result remains accurate, but it is time to change the battery.
2. The "
" symbol appears with E-b, Error and low:
The power is not enough to do a test. Please change the battery immediately.
10.2 REPLACING THE BATTERY
To replace the batteries (m), make sure the meter is turned off.
1. Press the edge of the battery cover and lift it up to remove.
2. Remove the old batteries and replace with two 1.5V AAA size alkaline batteries.
3. Close the battery cover. If the batteries are inserted correctly, you will hear a "beep"
afterwards.
18
BewellConnect - BW-GL1 - User Manual - 022017v1
All manuals and user guides at all-guides.com
OR
NOTE:
• Replacing the batteries does not affect the test results stored in the memory.
• As with all small batteries, these batteries should be kept away from children. If swallowed,
promptly seek medical assistance.
• Batteries might leak chemicals if unused for a long time. Remove the batteries if you are not
going to use the device for an extended period (i.e., 3 months or more).
• Properly dispose of the batteries according to your local environmental regulations
10.3 CARING FOR YOUR METER
CLEANING
1. To clean the meter exterior, wipe it with a cloth moistened with tap water or a mild
cleaning agent, then dry the device with a soft dry cloth. DO NOT rinse with water.
2. DO NOT use organic solvents to clean the meter.
METER STORAGE
• Storage conditions: -4°F to 140°F / -20°C to 60°C , below 95% relative humidity.
• Always store or transport the meter in its original storage case.
• Avoid dropping and heavy impact.
• Avoid direct sunlight and high humidity.
METER DISPOSAL
The used meter should be treated as a contaminated device that may carry a risk of
infection during measurement. The batteries in this used meter should be removed and
the meter should be disposed in accordance with local regulations.
10.4. CARING FOR YOUR TEST STRIPS
• Storage conditions: 35.6°F to 89.6°F / 2°C to 32°C, below 85% relative humidity. DO NOT freeze.
• Store your test strips in their original vial only. Do not transfer to another container.
• Store test strip packages in a cool dry place. Keep away from direct sunlight and heat.
• After removing a test strip from the vial, immediately close the vial cap tightly.
• Touch the test strip with clean and dry hands. Use each test strip immediately after
removing it from the vial.
• Write the opening date on the vial label when you first opened it. Discard remaining test
strips after 6 months.
• Do not use test strips beyond the expiry date. This may cause inaccurate results.
• Do not bend, cut, or alter a test strip in any way.
• Keep the strip vial away from children since the cap and the test strip may be a choking
hazard. If swallowed, promptly see a doctor for help.
For further information, please refer to the test strip package insert.
BewellConnect - BW-GL1 - User Manual - 022017v1
19