Personnel responsible for pump operation and maintenance must be
well-trained and must be aware of the accident prevention rules.
In addition to the requirements listed below, please refer to the
paragraph "Safety rules".
WARNING!
Introduce air through the inlet port before every maintenance operation.
WARNING!
If you have to perform maintenance on the pump after a considerable time in
operation, leave it to cool as temperature of the outer surface may be in
excess of 80 °C.
WARNING!
Always disconnect the power supply to the pump before starting maintenance
work. Place a special warning signs over the power supply breaker switch:
MACHINE UNDERGOING MAINTENANCE - DO NOT POWER ON. When
finished, remove the safety warning.
In order to keep the pump operating at a high efficiency level, it is
mandatory to follow all periodical service points listed in the table below.
However, more frequent service operations may be necessary
depending on what the pump is used for (suction of condensable
vapors, suction polluting substances).
For such cases, only direct experience can indicate the correct
service frequency needed. The exhausted oil and the replaced spare
parts must be considered as special waste products and handled
according to the local regulations in the Country of use.
MS120 Single Stage Rotary Vane Pumps User Manual/87-900-150-01 (C.00)
Instructions for Use

Maintenance

151/193
4