Guidance and manufacturer's declaration – electromagnetic emissions
The Rudolf Riester GmbH 222B is intended for use in the electromagnetic
environment specified below. The customer or the user of the The Rudolf Riester
GmbH 222B should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
N/A
CISPR 11
Harmonic emissions
N/A
IEC 6100-3-2
Voltage fluctuations/
N/A
flicker emissions IEC
61000-3-3
ri-cardio
32
Electromagnetic environment – guidance
The Rudolf Riester GmbH 222B uses
RF energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
The Rudolf Riester GmbH 222B uses
batteries only and is not connected to
mains.
Guidance and manufacturer's declaration – electromagnetic immunity
The Rudolf Riester GmbH 222B is intended for use in the electromagnetic
environment specified below. The customer or the user of the The Rudolf Riester
GmbH 222B should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Electrostatic discharge
±6 kV contact
(ESD) IEC 61000-4-2
±8 kV air
Electrical fast
N/A
transient/burst IEC
61000-4-4
Surge IEC 61000-4-5
N/A
Voltage dips, short
N/A
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency (50/60
3 A/m
Hz) magnetic field IEC
61000-4-8
ri-cardio
Compliance
Electromagnetic en-
level
vironment – guidance
±6 kV contact
Floors should be
±8 kV air
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
N/A
N/A
N/A
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical commercial or
hospital environment.
33