electrical stimulation may affect seizure
threshold.
Caution should be observed when
using the device at the same time as
being
connected
equipment with body worn electrode
pads. It may interfere with the signals
being monitored.
Caution should be used following
recent surgical procedures. Stimulation
may disrupt the healing process.
Caution Simultaneous connection to
high frequency surgical equipment may
result in burns and damage to the
stimulator.
Caution Strong electromagnetic fields
(electro-surgery/ microwave cookers/
mobile phones) may affect the correct
operation of this unit. If it appears to
behave unusually, move it away from
these devices.
Caution Do not permit use by children
unable to understand the instructions or
persons with cognitive disabilities, i.e.;
Alzheimer's disease or dementia.
Caution Keep away from children
under 5 years of age. Long lead wires -
risk of strangulation in infants.
Caution should be observed when
using the Allura at high strength
settings. Prolonged use at high settings
may cause muscle injury or tissue
inflammation. The Allura is capable of
delivering outputs in excess of 10 mA
(RMS) or 10 V averaged over any
period of 1 sec.
Caution Not intended for use in an
oxygen rich environment.
Caution Not intended for use in
conjunction
with
anaesthetics or flammable agents.
to
monitoring
flammable
Caution The patient is an intended
operator.
Caution Do not service and maintain
the device while in use.
Caution Maintenance and all repairs
should only be carried out by an
authorized agency. The manufacturer
will not be held responsible for the
results of maintenance or repairs by
unauthorized persons.
If necessary, we will provide circuit
diagrams, component part lists or other
information that will assist authorized
service personnel to repair the device.
Caution The operator should not touch
the patient at the same time when
touching the battery output.
Caution Keep away from pets and
pests.
DO NOT PLACE ELECTRODE PADS:
• On skin, which does not have normal
sensation. If the skin is numb too
great a strength may be used, which
could result in skin inflammation.
• On broken skin. The electrode pads
could encourage infection.
• On the front of the neck. This could
cause the airway to close, giving
breathing problems.
sudden drop in blood pressure
(vasovagal response).
• Over the eyes. May affect eyesight
or cause headaches.
• Across the front of the head. Effect
on patients who have had strokes or
seizures is not known. May affect
your sense of balance. The effects of
stimulation
on
unknown.
8
May
cause
the
brain are