Z-Lock:
Utilized to prevent inadvertent re-zeroing of
the gauge.
1. In Menu mode, press
"ZLOCK" appears.
|
|
2. Press
3. Press
,
OFF" (default).
|
|
4. Press
will briefl y show "DONE") and return to
Measurement mode.
Note: If Z-Lock is activated, the gauge will
display "ZLOCK" if zero is attempted.
,
to select.
to select "L ON" or "L
to select value (display
Re-Calibration:
Provides the user the ability to fi eld calibrate
the product. Original factory calibration is
permanently retained in memory and can be
until
recalled at any time.
1. In Menu mode press
"RECAL" appears; proceed to press
|
|
2. Display will indicate "FACT" (factory)
"FIELD" or "NEW". Press
to scroll.
3. Pressing
is displayed will restore values of factory
calibration.
4. Pressing
displayed will restore values from latest
fi eld calibration.
5. Pressing "Enter" key
"NEW" appears will enter the Recalibrate
mode.
6. When "NEW" is displayed, press
upon release, the display will fl ash "OK/
APPLY/ REF/ PSI/ THEN/ PRESS/ ENTER/
TO/ START/ OR/ OTHER/ TO/ ABORT".
Numeric display will read .00000. Vent
sensor to atmospheric pressure, press
|
|
count down from 6 seconds then briefl y
display CAL then automatically go to next
step.
7. Display flashes "APPLY/ REF/ PSI/
THEN/ PRESS/ ENTER/ TO/ START/
OR/ OTHER/ TO/ ABORT"; apply the full
scale pressure in units of psi indicated in
numeric display to gauge. Press
display will show "WAIT" and count
down from 6 seconds then briefl y display
CAL then automatically go to next step.
11
,
.
|
|
while "FACT" (factory)
|
|
when "FIELD" is
|
|
display will show "WAIT" and
until
,
when
|
|
 ;
|
|