Light:
Determines how long the backlight will remain
ON after any key is pressed in Measurement or
Menu modes (Note: The timer is reset with any
key being pressed.)
1. While in Menu mode, press
until "LIGHT" is displayed.
|
|
2. Press
for timer value to appear;
"ON" refers to the backlight remaining
ON at all times unit is powered ON,
"PRESS" designates that the backlight
is switched on / off by briefl y pressing
the backlight
indicates the backlight will never be
illuminated. Selecting time values, "1
MIN" (default), "5 MIN", "20 MIN", will
activate the backlight symbol on the LCD
display for the designated minutes.
|
|
3. Press
(display will briefl y show "DONE") and
return to Measurement mode.
Note: In "PRESS" mode; factory has set 1 hour
timer to save battery.
Feature: Backlight indicator will blink 10
seconds prior to light shut off .
,
whereas, "OFF"
menu
to select backlight time
Update:
Utilized to select the rate at which the displayed
pressure value is updated on the screen.
This function is used when rapid changes in
pressure cause "fl utter" in the display values;
longer intervals will reduce the update rate and
"average" the readings on such applications.
1. In Menu mode, press
"UPDAT" appears.
2. Press
3. Press
"1 SEC" (default), "500 MSEC", or "250
MSEC".
4. Press
will briefl y show "DONE") and return to
Measurement mode.
Note: Changing value to anything other than
1 sec may cause a slight zero off set, and
it is recommended that the gauge be fully
vented and re-zeroed before taking accurate
readings. Also, battery life will be reduced by
use of an update rate faster than 1 SEC.
10
,
|
|
to select.
,
to select values for
|
|
to select value (display
until