 Press it once to turn on the Refrigerant leak de-
tector, the display will illuminate with flash, for 45
seconds to eat up the sensor. After this period the
unit can be used.
Auto reset & Reset function key:
When the Auto Reset function is turned ON, the meter
will monitor background status and fine tune itself.
When Reset LED light is on, it indicates it is in ON
mode.
 Press Reset button and hold for 2 seconds the
Reset light will turn off and Auto Reset function is in
OFF mode.
When the Reset light is off, it indicates the Reset func-
tion is in manual mode.
 Press the Reset button once to enable manual
Reset function.
Verify the condition of the unit and sensor:
 Set the sensitivity level to "Hi".
 Open the leak check bottle cover and slowly move it
closer to snake tube nozzle.
If the indication moves up to high from low then we
should move the check bottle away and the LED should
go off again. This shows that the unit is under working
condition.
If the unit does not perform as we expect, bring the unit
for maintenance at your local sales office.
Enter the measuring mode:
 Place the tip of the leak-detector probe as close
as possible to the site of the suspected leak. Try
to position the probe within 1/4 inch (6 mm) of the
possible leak source.
 Slowly move the probe past each possible leakage
point.
 When the instrument detects a leak source, the
audible tone will alarm. Additionally, the visual
indicators will light from left to right, Green LED then
Orange LED then Red LED (highest concentration)
as increasing of level indicate that the location is
close to the source.