Section 1 Introduction
Brooks
Ar-Mite
®
1-3 Optional Equipment
1-4
MT3750C
TM
Standard Needle Valves
The standard needle or cartridge valves can be supplied integrally
mounted to the inlet fitting of the instrument. The optional NRS
valve, also integrally mounted to the inlet fitting, provides a greater number
of turns affording precision control. For more details on the needle or
cartridge valves go to our website: BrooksInstrument.com,
select Documentation,Precision Valves and Flow Controllers,
Brooks-Line IV
Limit Switches
Inductive Alarm Switch
One or two electronic limit switches type SJ2-N can be installed in the
indicator housing to allow initiation of signaling or switching functions on a
preset flow value being reached. The SJ2-N limit switch operates as a slot
initiator that is inductively actuated by a cam mounted to the pointer. Any
flow value can be used for setting the limit value by sliding the switch along
the slot in the mounting plate for the initiators. Minimum setting distance
between two limit switches is approximately 50% of the scale range.
(Refer to Figure 1-1)
Power supply
Current consumption
Current consumption
Self inductance
Self capacitance
Max Temp:
Figure 1-1 Model MT3750C with Transmitter / Inductive Alarm
Installation and Operation Manual
, CART valves or NRS valves.
TM
8 Vdc (Max. 15.5 Vdc)
active area clear: > 3 mA
active area obscured: < 1 mA
29 μH
20 nF
158°F(70°C)
X-VA-MT3750C-eng
Part Number: 541B063AAG
October, 2013
needle
TM