5 | Assembly
5.1.2 Mounting the pyranometer
Required tools
– Slotted screwdriver, 8 mm
– Wrench, 8 mm
1
2
3
4
5
4 To insulate the pyranometer against the temperature of the mounting device, place the pyranometer on
the adjustable foot and the two leveling feet.
4 Position the pyranometer in such a way that the nuts are located at a distance of 2 to 3 mm from the
mounting device.
4 Make sure that the pyranometer is grounded.
4 Make sure that the pyranometer is not in the shade.
4 Align the signal line with the nearest pole to prevent electrical components from heating up.
28
65 mm
1
2x M5 x 80 mm screws
2
2x nylon insulating rings
3
2x Ø 5.2 mm
4
2x washers
5
2x nuts
SMP12 Smart Pyranometer | EN