The first and fundamental step in achieving good processing, long life and complete
use of its features is a correct, careful installation. Moreover, this precaution will avoid
the danger of physical injury or property damage, not to mention malfunctions and
damage to the machine. So, please follow the instructions in this chapter
scrupulously.
INSTALLATION MUST BE PERFORMED BY QUALIFIED
PERSONNEL.
M.O.COM C
USTOMER SUPPORT
ANY DOUBT OR SUPPLY FURTHER INFORMATION
T
HE UNIT IS SUPPLIED AS TESTED AND AFTER HAVING PASSED ALL THE
SCHEDULED FACTORY TESTS
Size and weight
– Height (total)
555 mm
– Length (total)
560 mm
– Depth
600 mm
– Overall weight
about 44 kg
Power supply
The electrical system to which the sterilizer will
be connected must be suitably dimensioned
based on the electrical characteristics of the
device. This information is shown on the back
of the machine.
-
Install the unit on a plane strong enough to support its weight (44 kg) and at a
height of at least 25 cm; if necessary adjust the feet to compensate any plane
inclination.
LEAVE ADEQUATE SPACE FOR VENTILATION (AT LEAST 10 CM
ON EACH SIDE) ALL AROUND THE UNIT, ESPECIALLY IN THE
BACK AND UPPER TO ALLOW ITS CORRECT VENTILATION.
-
Do not install the sterilizer near tubs, sinks or similar places, to avoid contact with
water or liquids. This could cause short circuits and/or potentially dangerous
situations for the operator;
-
Do not install the sterilizer in a place that is excessively humid or poorly ventilated;
-
Do not install the machine were there is gas or inflammable and/or explosive
vapors;
-
Install the unit so that the power cord and the inlet/outlet pipes are not bent or
pressed.
STEAM OUTLET FROM THE BACK.
INSTALL THE UNIT SO THAT THE BACK FACES A WALL, NOT A
SPACE FOR STANDING OR PASSING BY.
WARNING
NOTE
. (
Z)
SEE APPENDIX
IS AVAILABLE TO CLEAR
.
.
555
600
WARNING
WARNING
560