Accuracy
Accuracies apply for 1 year 18°C to 26°C. Temperature coefficient outside these limits is
<0·1 x quoted range accuracy per °C. Warm-up time to rated accuracy is 1 hour.
DC Volts
Range
200mV
2V
20V
200V
1000V
Resistance (Ohms)
Range
200
2k
20k
200k
2M
20M
Max open circuit voltage 3·5V.
Audible continuity check available on all ranges.
AC Volts (True RMS)
Range
200mV
2V
20V
200V
750V
Accuracy specifications apply for readings between 10% and 100% of full scale.
Additional error at crest factor = 3 typically 0·7%.
Input impedance 1MΩ || <100 pF any range.
Max input 750V rms, 1kV peak, any range.
DC Amps
10A, up to 2000mA
10A, above 2000mA
2
Accuracy
0·017% + 3d
0·012% + 3d
0·019% + 3d
0·019% + 3d
0·019% + 3d
Accuracy
0·025% + 4d
0·019% + 3d
0·019% + 3d
0·019% + 3d
0·022% + 3d
0·07% + 3d
45Hz - 5kHz
0·2% + 150d
0·2% + 150d
0·2% + 150d
0·2% + 150d
0·2% + 150d
Range
Accuracy
200µA
0·08% + 12d
2mA
0·08% + 12d
20mA
0·08% + 12d
200mA
0·08% + 12d
0·08% + 12d
0·12% + 12d
Input Impedance
10MΩ or >1GΩ
10MΩ
Max. Measuring Current
1·5mA
1·5mA
0·4mA
40µA
4µA
400nA
Accuracy
5kHz - 10kHz
10kHz-20kHz
0·2% + 250d
3% + 500d
0·2% + 250d
0·5% + 300d
0·2% + 250d
0·5% + 300d
0·2% + 250d
0·5% + 500d
-
Voltage Burden
300 mV max
650mV max
Specification
Max Input
1kV DC or AC
Peak any range
Max. Input
300V DC or RMS
continuous any
Ohms input any
range
20kHz-50kHz
-
3% + 1500D
2% + 1000D
-
-
-
Max Input
1A, 300V
fuse protected
10A, 300V
fuse protected