Storing the devices
Store the transmitter, charger, and tester in a clean, dry location at room temperature. If
the transmitter is not in use, you must charge the transmitter at least once every 60 days.
Although not required, you may store the transmitter on the charger. If you are storing the
transmitter on the charger, you must disconnect and reconnect the charger and the
transmitter at least once every 60 days.
Disposal
Discard the transmitter according to local regulations for battery disposal.
Specifications
Biocompatibility
Applied parts
Operating conditions
Storage conditions
Battery life
Transmitter frequen-
cy
Maximum output
power (EIRP)
Transmitter: Complies with EN ISO 10993-1
Transmitter
Sensor
Transmitter temperature: -5 °C to 45 °C (23 °F to 113 °F)
Caution: When operating the transmitter on a tester in air temperatures
greater than 41 °C (106 °F), the temperature of the transmitter may exceed
43 °C (109 °F)
Transmitter relative humidity: 5% to 95% with no condensation
Transmitter pressure: 61.36 kPa to 106.17 kPa (8.9 psi to 15.4 psi)
Charger temperature: 10 °C to 40 °C (50 °F to 104 °F)
Charger relative humidity: 30% to 75% with no condensation
Transmitter temperature: -25 °C to 55 °C (-13 °F to 131 °F)
Transmitter relative humidity: 10% to 100% with condensation
Transmitter pressure: 61.36 kPa to 106.17 kPa (8.9 psi to 15.4 psi)
Charger temperature: -10 °C to 50 °C (14 °F to 122 °F)
Charger relative humidity: 10% to 95% with no condensation
Uses one new AAA battery to charge the transmitter.
2.4 GHz, 2M65G1D modulation, less than 1mW ERP
-0.63 dBm