Error Messages; Display Description - GERATHERM GT-101 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
DE
EN

ERROR MESSAGES

Display
FR
IT
Description
The measured temperature is
too high
H appears if the reading is
above 42.2 °C (108 °F) in body
mode or 100 °C (212 °F) in
object mode.
The measured temperature
is too low
L appears if the reading is
below 34 °C (93.2 °F) in body
mode or 0 °C (32 °F) in object
mode.
Ambient temperature is
too high
H appears in connection with
if the ambient temperature
is above 40 °C (104 °F).
Ambient temperature is
too low
L appears in connection with
if the ambient temperature
is below 16 °C (60.8 °F) in
body mode or 5 °C (41 °F) in
object mode.
Error indication on the
display
3 short signal beeps will sound.
Blank display
Check the position or polarity of
the battery.
Flat battery indication
Insert new batteries.