NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re-
flection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
a
radios, amateur radio,AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be consi-
dered. If the measured field strength in the location in which the Blood pressure monitor Medel Control is used exceeds
the applicable RF compliance level above, the Blood pressure monitor Medel Control should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating
the Blood pressure monitor Medel Control.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
95142_MEDELcontrol_userMANUAL_REV.00_02.03.2016 TRAD.indd 147
All manuals and user guides at all-guides.com
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya should be less than the compliance
level in each frequency rangeb. Interference may occur in the vicinity
of equipment marked with the following symbol: