10. In the Edit an application window, click the Configuration tab
and check/edit the parameters of the sample changer used.
- Type of sample changer (SAC950),
- Number of rinses. Usually 2 static rinses are performed of 20
seconds each.
You can also define one or several (up to 9) dynamic rinses. A
dynamic rinse is a combination of a shower rinse of the electrodes
and an emptying operation of the beaker (consult the SAC950
User's Guide, part no. D21T085 or the on-line help of the
TitraMaster 85 PC software).
Important:
If you define a dynamic rinse, you must also define one or two
static rinses as the TTL 5 V triggering the dynamic pump module
can only be activated at a start of a static rinse and not at a start
of a dynamic rinse.
Do not define a dynamic rinse and do not change the number/
duration of static rinses (2 rinses of 20 s each by default) without
having read chapter 3.
Do not change the other settings of the Configuration window.