Inficon TEK-Mate 705-202-G Serie Mode D'emploi page 8

Détecteur de fuites de fluide frigorigène
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PROBLEM
4. False alarms -
TEK-Mate alarms
when the probe is
moved or bumped.
Return Authorization Procedure
All defective TEK-Mates should be returned to your wholesaler for warranty
evaluation. If you have any questions, please contact INFICON at
+49 221 56788-660 (servicetools.europe@inficon.com).
NOTE: Do not return your defective unit directly to the factory without first
contacting your wholesaler.
CAUSE
4a. Sensor leads are
bent.
4b. Moisture was
absorbed by the
sensor during a long
period without use.
8
REMEDY
4a. Remove the sensor and
inspect the leads. Straighten
the leads with needle nose
pliers, if necessary, and
reinstall the sensor.
4b. Run TEK-Mate for at least
20 minutes. The absorption of
moisture does not affect the
life or sensitivity of the sensor.